Supercritical Fluid Clean-Up of Environmental Samples for the Analysis of Polycyclic Aromatic Hydrocarbons Using Time-of-Flight Secondary Ion Mass Spectrometry

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 参考文献18件

この論文をさがす

抄録

A novel sample-pretreatment method for time-of-flight secondary ion mass spectrometry (TOF-SIMS) was developed using supercritical fluid extraction (SFE). In SFE, the extraction efficiency of a certain organic matter is controlled by the pressure and temperature of supercritical CO2. Two-step SFE (1st step at 10 Mpa, 40°C; 2nd step at 30 MPa, 120°C) was applied to diesel exhaust particles containing many kinds of n-alkanes and aromatic species. n-Alkanes and polycyclic aromatic hydrocarbons (PAHs) were extracted in the 1st and 2nd steps, respectively. This selectivity was utilized for the sample preparation of TOF-SIMS analysis. Diesel exhaust particles after the 1st step of extraction were analyzed with TOF-SIMS, aiming at PAHs as analytical targets. The obtained spectrum was simplified, and mass peaks of individual PAHs were easily assigned, because unwanted compounds, like n-alkanes, were selectively removed by SFE. Furthermore, a simple calculation elucidated the outline of the spectrum.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 20 (10), 1379-1382, 2004

    社団法人 日本分析化学会

参考文献 (18)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ