Hydrogen Ion-Selective Poly(vinyl chloride) Membrane Electrode Based on a Calix[4]arene.

DOI PDF Web Site Web Site Web Site View 1 Remaining Hide 1 Citations 72 References

Search this article

Abstract

A hydrogen ion-selective poly(vinyl chloride) membrane electrode was constructed using 5,11,17,23-tetra-tert-butyl-25,26,27,28-tetracyanomethoxycalix[4]arene as a neutral carrier. The electrode showed an apparent Nernstian response in the 2 - 11.5 pH range with a slope of 54.0±0.2 mV/pH at 20±1°C. This electrode showed a rapid response of the emf to changes in the pH, high ion selectivity with respect to lithium, sodium and potassium, and characteristics similar to those reported for the conventional pH glass membrane electrode. It can be used as a potentiometric indicator electrode in hydrofluoric acid solutions. The effects of iodide, thiocyanate, perchlorate and bromide on the characteristics of the electrode were also considered.

Journal

  • Analytical Sciences

    Analytical Sciences 19 (2), 217-221, 2003

    The Japan Society for Analytical Chemistry

Citations (1)*help

See more

References(72)*help

See more