Capillary Electrophoresis-Chemiluminescence Detection System Equipped with a Consecutive Sample-Injection Device

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献2件 参考文献8件 オープンアクセス
  • TSUKAGOSHI Kazuhiko
    Department of Chemical Engineering and Materials Science, Faculty of Engineering, Doshisha University
  • NAKAHAMA Koji
    Department of Chemical Engineering and Materials Science, Faculty of Engineering, Doshisha University
  • NAKAJIMA Riichiro
    Department of Chemical Engineering and Materials Science, Faculty of Engineering, Doshisha University

この論文をさがす

説明

We have developed a consecutive sample-injection device for capillary electrophoresis, which comprises one four-way cock, two syringe pumps, and an interface part taking advantage of two three-way Teflon joints. Sample introduction into the capillary is made hydrodynamically by pressure, caused by the flow of the sample solution at the tip of the capillary inlet. We combined the injection device with a capillary electrophoresis-chemiluminescence detection system. A mixture solution of N-(4-aminobutyl)-N-ethylisoluminol, isoluminol isothiocyanate, and luminol was analyzed as a model sample by the present system. The sample solution was consecutively injected and detected with about a 230 s interval. The present capillary electrophoresis-chemiluminescence detection system with the consecutive sample injection device features easy and rapid operation, an inexpensive apparatus, high sensitivity, as well as consecutive analysis.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 20 (2), 379-381, 2004

    社団法人 日本分析化学会

被引用文献 (2)*注記

もっと見る

参考文献 (8)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘