Chemiluminescence Detection of Glucose Using Ce(IV) Oxidation in a Batch System

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 参考文献35件
  • NAGAI Megumi
    Separation Science Division, Faculty of Science, Okayama University of Science
  • MIYAHARA Wakayo
    Separation Science Division, Faculty of Science, Okayama University of Science
  • SAGAWA Honami
    Separation Science Division, Faculty of Science, Okayama University of Science
  • YAMAZAKI Shigeo
    Separation Science Division, Faculty of Science, Okayama University of Science
  • SAITO Keiitsu
    Division of Natural Environment and Chemistry, Faculty of Human Development, Kobe University

この論文をさがす

抄録

A chemiluminescence method for the determination of glucose (GC) using a batch system made in our laboratory is proposed. The method is based on measures of the chemiluminesence intensity from a batch of sample GC solutions following the injection of an (NH4)2[Ce(NO3)6]–[Ru(bpy)3]Cl2 solution. The effect of the Ru complex concentration on chemiluminescence was studied. The detection limit was 0.5 × 10−8 M, and the linear calibration curve reached 10−7 M. The Ru complex was selected on the basis of its desirable performance compared to those of other photosensitizers. However, flow-injection analysis (FIA) using the Ce-Ru complex reagent was not applicable for the detection of GC.