Determination of Inorganic Anions by Capillary Ion-exchange Chromatography Using Polyethylenimine-coated Octadecyl-bonded Phases

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献1件 参考文献62件

この論文をさがす

抄録

The utility of a cationic polymer polyethylenimines (PEI) for coating some commercially available silica-based stationary phases packed in capillaries of 0.32 mm i.d. and 10 cm long was studied for the separation and direct UV detection of inorganic anions (iodate, bromate, nitrite, bromide, and nitrate ions). With a super-endcapped octadecylated silica stationary phase, which yielded the best separation in terms of resolution and retention time of individual anions, the effect of the pH and composition of the eluent on the elution behavior of individual anions was studied. The relative standard deviations of the retention time, peak area and peak height for 60 successive injections (running time of more than 10 h) were not more than 1.32, 2.30 and 1.94, respectively, with the exceptions of the peak area and peak height of the bromate ion. The detection limits at 210 nm ranged from 1 to 7 ppm. This method was successfully applied to determine concentrations of nitrate ions present in pills and beverage samples.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 28 (6), 625-629, 2012

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

参考文献 (62)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ