レムカウンターによる電子ライナック周辺のパルス混成放射線場における中性子線量当量評価

DOI 被引用文献1件 オープンアクセス

書誌事項

タイトル別名
  • Evaluation of neutron dose equivalent in pulsed and mixed radiation field around electron linear accelerator with rem-counter.

この論文をさがす

説明

Two Important problems about commonly-used neutron rem-counter have been discussed in evaluating the neutron dose equivalent in pulsed and mixed radiation fields around an electron linear accelerator. One is the overestimate of dose equivalent due to false pulses generated by intense burst X-rays. In order to eliminate the false pulses, three techniques based on (1) anti-coincidence counting, (2) pulse height discrimination, and (3) correction factor have been employed. It was pointed out that the first technique was most effective and the third was most convenient for practical applications. The other problem is the underestimate of dose equivalent due to counting loss in a high-fluence-rate field. It was confirmed that the time variation of counting rate was an important factor for dead-time correction in pulsed radiation fields. A comparison between dose equivalent values evaluated with rem-counter and those with Indium activation detector has also been carried out. A validity of proposed correction methods was confirmed from the experimental results of the proportionality of the corrected dose equivalent values to the relative activity of Indium detectors.

収録刊行物

  • 保健物理

    保健物理 20 (4), 371-378, 1985

    一般社団法人日本保健物理学会

被引用文献 (1)*注記

もっと見る

詳細情報 詳細情報について

  • CRID
    1390001204288181888
  • NII論文ID
    130000793013
  • NII書誌ID
    AN00228458
  • DOI
    10.5453/jhps.20.371
  • COI
    1:CAS:528:DyaL2sXjvF2jsg%3D%3D
  • ISSN
    18847560
    03676110
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • Crossref
    • CiNii Articles
    • OpenAIRE
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ