イオン選択性電極に対する濃度直線指示変換器の試作

書誌事項

タイトル別名
  • Linear ionic concentration meter with ion selective electrodes
  • イオン センタクセイ デンキョク ニ タイスル ノウド チョクセン シジ ヘン

この論文をさがす

説明

In order to avoid the usual difficulties such as the elaborate calculations or the use of diagrams which are encountered in the quantitative analysis by means of the ion selective electrodes, a linear ionic concentration meter has been built which utilized the anti-log conversion function of ion selective electrode. The meter possesses two abilities, one, arbitrary setting of Nernst slope and the other, arbitrary setting of reference point in concentration scale. These two functions are called as Slope and Shifter, respectively.<BR>In the system of AgI-I, good linear relation was obtained over a hundred times range of concentration. Another application of the meter to the combined procedure of Gran's plots and standard addition method also showed good results.<BR>The process is appropriate for the combined method and provides an extremely simple procedure. Discussion is given about the relative errors due to the setting errors of both the Slope and Shifter.

収録刊行物

  • 分析化学

    分析化学 24 (2), 101-106, 1975

    公益社団法人 日本分析化学会

詳細情報 詳細情報について

問題の指摘

ページトップへ