書誌事項
- タイトル別名
-
- Biological Temperature Measurement Using a Microthermocouple Probe
- マイクロ ネツデンツイ プローブ オ モチイタ セイタイ オンド ケイソク
この論文をさがす
説明
A responsive microthermocouple probe with a time constant of 2-3 ms was constructed using a fabrication method for glass microelectrodes as widely utilized in neurobiology. The probe is composed of a constantan wire of 25 μm in diameter and a vapor-deposited copper film of 1.8 μm in thickness. These two components come into contact at the tip of a glass pipette, forming a sensing junction of approximately 17 μm in diameter. The thermoelectromotive force of 40.2 μV/°C is very close to the standard value of the International Electrotechnical Commission and linearly dependent on temperature over the range normally encountered in biological measurement. The microthermocouple probe was tested by insertion into a dorsal longitudinal muscle of a silkworm moth. The core temperature was found to rise rapidly by 10-11 °C after pheromonal stimulation, returning to the original temperature at rest in 15-20 min. Moreover, the probe was used successfully to measure the temperature variation in the thorax of a cockroach without damage to the tissue.
収録刊行物
-
- 電気学会論文誌E(センサ・マイクロマシン部門誌)
-
電気学会論文誌E(センサ・マイクロマシン部門誌) 122 (12), 567-572, 2002
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001204461420544
-
- NII論文ID
- 10010455268
-
- NII書誌ID
- AN1052634X
-
- ISSN
- 13475525
- 13418939
-
- NDL書誌ID
- 6368851
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可