Development and Evaluation of Microfluidic Device for Mn Ion Quantification in Ocean Environments
-
- Fukuba Tatsuhiro
- Institute of Indutrial Science, The Univerity of Tokyo
-
- Provin Christophe
- Institute of Indutrial Science, The Univerity of Tokyo
-
- Okamura Kei
- Center for Advanced Marine Core Research, Kochi University
-
- Fujii Teruo
- Institute of Indutrial Science, The Univerity of Tokyo
この論文をさがす
抄録
This paper presents design, fabrication and evaluation of a polydimethylsiloxane (PDMS) —glass microfluidic device for quantitative determination of manganese ion (Mn(II)) in oceanic environments. The microfluidic device can perform luminol-hydrogen peroxide chemiluminescence (CL) reaction in a flow-through manner for highly sensitive Mn (II) quantification. A chelate resin column was integrated in the microfluidic device for removal of coexisting metal ion such as Fe(II). The developed device could quantify Mn(II) in nM level under coexistence of μM level of Fe(II). The microfluidic device developed here will be used as a core element of a totally integrated in situ chemical analyzer that we have been developing for oceanography applications.