書誌事項
- タイトル別名
-
- Review of Microfluidic Devices for On-Chip Chemical Sensing
- オンチップ カガク センシング ノ タメ ノ マイクロフルイディックデバイス ニ カンスル レビュー
この論文をさがす
説明
Microfluidic techniques can be a basis to realize user-friendly microanalytical devices of high performance. Required operations include injection of solutions to reaction chambers, volume measurement, and mixing of solutions. For this purpose, the use of surface tension is attractive. In hydrophobic flow channels, solutions can be moved stably in the form of plugs. Otherwise, solutions can be transported by capillary action in a hydrophilic flow channel and stopped at a hydrophobic valve. The valve can be opened autonomously by switching the mixed potential by wetting a zinc electrode formed in a controlling flow channel. In this review paper, we introduce some of our microfluidic devices particularly focusing on devices that work based on surface tension.
収録刊行物
-
- 電気学会論文誌E(センサ・マイクロマシン部門誌)
-
電気学会論文誌E(センサ・マイクロマシン部門誌) 136 (6), 244-249, 2016
一般社団法人 電気学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204462518784
-
- NII論文ID
- 130005154787
-
- NII書誌ID
- AN1052634X
-
- ISSN
- 13475525
- 13418939
-
- NDL書誌ID
- 027489693
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可