書誌事項
- タイトル別名
-
- Intraluminal MRI Probe Using Small Size Variable Capacitor
- コガタ カヘン ヨウリョウ オ モチイタ タイコウ ナイ MRI プローブ ノ カイハツ
この論文をさがす
抄録
An intraluminal magnetic resonance imaging (MRI) probe holds promise to achieve a higher resolution image of a small pathological lesion than conventional external probes. Such an intraluminal probe requires precise resonant frequency tuning to 63.865 MHz and impedance matching to 50 Ω, if used for instance in a 1.5-T scanner. Tuning and matching have to be performed after inserted into the subject body as electric permittivity varies according to the tissue location. Thus, in this study, a novel intraluminal MRI probe which can be tuned and matched remotely by voltage-controlled, MOSFET-type variable capacitors was developed and fabricated. The imaging capability and signal-to-noise ratio (SNR) was evaluated in an agar phantom.
収録刊行物
-
- 電気学会論文誌E(センサ・マイクロマシン部門誌)
-
電気学会論文誌E(センサ・マイクロマシン部門誌) 136 (5), 153-159, 2016
一般社団法人 電気学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204462580480
-
- NII論文ID
- 130005149049
-
- NII書誌ID
- AN1052634X
-
- ISSN
- 13475525
- 13418939
-
- NDL書誌ID
- 027423681
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可