- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Investigation of optical-axis adjusting system for a slit-scan flow cytometer
-
- Tokushima Hisao
- Department of Advanced System Control Engineering, Graduate School of Science and Engineering, Saga University
-
- Dozono Hiroshi
- Department of Advanced System Control Engineering, Graduate School of Science and Engineering, Saga University
-
- Hara Shigeomi
- Department of Electronic Engineering and Electronics, Faculty of Science and Engineering, Saga University
-
- Noguchi Yoshio
- Department of Advanced System Control Engineering, Graduate School of Science and Engineering, Saga University
Bibliographic Information
- Other Title
-
- スリットスキャンフローサイトメータの光軸調整に関する検討
Search this article
Description
<p>A flow cytometer is used for rapid analyses of cells or chromosomes. For an optical system of flow cytometer, it is important to adjust the center of illuminating laser beam to a jet-stream. Microspheres or leukocytes are used for focus adjustment. Adjusted best position is the location where fluorescence of standard samples becomes the brightest. Manual adjustment of the optical-axis is not only complicated but also inaccurate. A slit-scan flow cytometer measures the morphological information along the objects being analyzed. The laser beam size of our instrument at the tightly focused position has dimensions of 0.95 in the flow direction by 85.0 across with a depth of focus being only 1. At unfocused position, width of the slit becomes wider and intensity of the peak beam becomes weaker. To guarantee best focusing, we examined a method of opticalaxis adjustment for the slit-scan flow cytometer by controlling position of the jet-stream based on fluorescence intensities of microspheres passed. The computer controlled optical adjustment system ensures stable and reliable measurements by the slit-scan flow cytometer.</p>
Journal
-
- Cytometry Research
-
Cytometry Research 14 (2), 33-37, 2004
Japan Cytometry Society
- Tweet
Details 詳細情報について
-
- CRID
- 1390001204472289408
-
- NII Article ID
- 10026899650
-
- NII Book ID
- AN10356261
-
- ISSN
- 24240664
- 09166920
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed