書誌事項
- タイトル別名
-
- Studies on the Practical Application of Ion Electrode Method. VII. Fundamental Analytical Conditions in the Determination of Chloride Ion
この論文をさがす
説明
In the continuous determination of chloride ion in a large number of sample solutions, concentrations of which varied as in river and sea water, it was confirmed that the ion-selective electrode method was much better than the conventional Mohr's method in the simplification of the procedure, because the calibration curve of the former method was almost linear in such a wide range of chloride ion concentrations as 10-100000 ppm. The ion-selective electrode method was interfered by a sulfide ion, which, however, could be eliminated by the addition of lead acetate solution. With this method, determination of 100 ppm chloride ion solution was not affected by cupric ion less than 10 ppm and mercuric ion less than 0.1 ppm.
収録刊行物
-
- 衛生化学
-
衛生化学 22 (5), 286-290, 1976
公益社団法人 日本薬学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204497619072
-
- NII論文ID
- 110003642596
-
- NII書誌ID
- AN00021540
-
- ISSN
- 0013273X
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可