Gas Chromatographic Determination of a New Fungicide, Denmert<sup>®</sup>

  • IRIE Shigeyuki
    Institute for biological Science, Sumitomo Chemical Co., Ltd.
  • YUKI Fusao
    Institute for biological Science, Sumitomo Chemical Co., Ltd.
  • MURANO Atsushi
    Institute for biological Science, Sumitomo Chemical Co., Ltd.

Bibliographic Information

Other Title
  • ガスクロマトグラフ法による新殺菌剤デンマートの分析
  • ガス クロマトグラフホウ ニヨル シン サッキンザイ デンマート ノ ブンセキ

Search this article

Description

A quantitative method for the determination of S-n-butyl S'-p-tert-butylbenzyl N-3-pyridyldithiocarbonimidate (Denmert_??_) and its formulations was developed by gas chromatography (GC). Denmert was separated from its m-isomer and impurities by an apparatus equipped with a hydrogen flame ionization detector (FID) and determined by using 1, 3, 5-triphenylbenzene as an internal standard, under the following GC conditions; gas chromatograph: Shimadzu GC-5 APTF with an FID; column: a glass column, 3mm i.d. and 0.5m in length, packed with 2% PEG 20M/Chromosorb W (AW, DMCS, 60_??_80 mesh); column temp.: 220°C; injection port and detector temp.: 270°C; carrier gas: N2, 80ml/min. Moreover, most of impurities in technical products of Denmert were identified by a mass spectrometer combined with GC and determined by programmed temperature GC using the same column.

Journal

  • Nippon Nōgeikagaku Kaishi

    Nippon Nōgeikagaku Kaishi 51 (5), 331-333, 1977

    Japan Society for Bioscience, Biotechnology, and Agrochemistry

Details 詳細情報について

Report a problem

Back to top