A closed pentane trap for separation of SO2 from CO2 for precise .DELTA.18O and .DELTA.34S measurements
-
- Kusakabe Minoru
- Institute for Study of the Earth's Interior, Okayama University
Bibliographic Information
- Other Title
-
- A closed pentane trap for separation of SO2 from CO2 for precise δ18O and δ34S measurements
- closed pentane trap for separation of SO2 from CO2 for precise デルタ 18O and デルタ 34S measurements
Search this article
Description
Some practical aspects of a closed n-pentane trap used for separation of SO2 from CO2 and vice versa for precise δ18O and δ34S measurements are described. It is a cold finger sealed inside a large-diameter glass tube that is filled with n-pentane. This trap has several advantages (1) complete separation of the two gases within a few of minutes, (2) no fire risk because the inflammable pentane is present in a closed system, (3) low cost because the pentane in the trap will never evaporate, and (4) ease of use.
Journal
-
- GEOCHEMICAL JOURNAL
-
GEOCHEMICAL JOURNAL 39 (3), 285-287, 2005
GEOCHEMICAL SOCIETY OF JAPAN
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001204549787136
-
- NII Article ID
- 10015661471
-
- NII Book ID
- AA00654975
-
- COI
- 1:CAS:528:DC%2BD2MXlt1CgsLw%3D
-
- ISSN
- 18805973
- 00167002
-
- NDL BIB ID
- 7323785
-
- Text Lang
- en
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed