書誌事項
- タイトル別名
-
- Development of the Skeletal Muscle Stimulator with Resonant Type Transmission Function.
- キョウシンガタ プログラム ツウシン キノウ オ ユウスル ウエコミガタ コッ
この論文をさがす
抄録
In order to investigate the technique of cardiac assistance using skeletal muscle (Cardiomyoplasty) we are developing an implantable skeletal muscle stimulator that responds to muscle fatigue. The stimulator has a resonant circuit to transmit internal information using the carrier waves from the external program head. The stimulator can also control the synchronization ratio automatically using the peak value of the impulse response from an acceleration sensor to estimate muscle fatigue. In this series of animal experiments we verified the functions of both data transmission by the resonant circuit and automatic control of the synchronization ratio. Our results suggest that the stimulator can be used to automatically control muscle stimulation in Dynamic Cardiomyoplasty.
収録刊行物
-
- 医用電子と生体工学
-
医用電子と生体工学 35 (1), 64-70, 1997
一般社団法人 日本生体医工学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001204557242624
-
- NII論文ID
- 10009790952
-
- NII書誌ID
- AN00016981
-
- ISSN
- 21855498
- 00213292
-
- HANDLE
- 10191/25892
-
- NDL書誌ID
- 4174280
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- IRDB
- NDL
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可