Frequency Difference between a Methane-Stabilized 3.39μm Laser and the Third Harmonic of the 10.2μm R (30) Transition of a CO<sub>2</sub> Laser

DOI

Bibliographic Information

Other Title
  • メタン安定化3.39μm He-Neレーザーと10.2μm R (30) CO<sub>2</sub>レーザーの周波数比較

Description

A high power He-Ne laser with a folded cavity was constructed to obtain a beat frequency with the third harmonic of the 10.2μm R (30) radiation of a CO2 laser. The He-Ne laser was offset locked to a methane-stabilized 3.39μm He-Ne laser. The beat frequency and the third harmonic of the CO2 laser are generated in a tungsten-nickel point contact diode. The measured frequency difference is 48 731.796±0.281MHz. This result agrees well with both the value obtained at the National Bureau of Standards and the value obtained at the National Physical Laboratory.

Journal

  • Oyo Buturi

    Oyo Buturi 50 (1), 29-34, 1981

    The Japan Society of Applied Physics

Details 詳細情報について

  • CRID
    1390001204595327488
  • NII Article ID
    130003590990
  • DOI
    10.11470/oubutsu1932.50.29
  • COI
    1:CAS:528:DyaL3MXlsVWms70%3D
  • ISSN
    21882290
    03698009