書誌事項
- タイトル別名
-
- A Measurement of Magnetocardiogram (MCG) in Multi Point using a Transmission Line Type Thin-Film Sensor
- デンソウ センロガタ ハクマク センサ ニ ヨル シン ジカイ ノ タテン ケイソク
この論文をさがす
抄録
We have developed a very sensitive thin film sensor using a transmission line. The sensor element consists of a coplanar line fabricated from a ceramic substrate and amorphous-CoNbZr film (4μm in thickness). The ceramic substrate with high permittivity enhances the phase change of the carrier, phase change around 7000 degree/Oe being achieved. A small phase change was detected using a Dual Mixer Time Difference (DMTD) method when a weak magnetic field are applied. We applied the sensor for magnetocardiogram (MCG) measurement using the phase detection technique. We succeeded in measuring the MCG signals in multi point, including a typical QRS wave and T wave, and this signals were found to roughly agree with already reported MCG data by SQUID.
収録刊行物
-
- 電気学会論文誌. A
-
電気学会論文誌. A 133 (6), 372-375, 2013
一般社団法人 電気学会