Reformation of Undershoot on Amplifier for Pulse Height Analysis
-
- KARIYA Komyo
- Department of Electricity, Faculty of Science and Engineering. Ritsumeikan University
-
- AOKI Yasusi
- Department of Electricity, Faculty of Science and Engineering. Ritsumeikan University
-
- NAKANISHI Tunehiko
- Department of Electricity, Faculty of Science and Engineering. Ritsumeikan University
-
- OKAMOTO Hiroshi
- Department of Electricity, Faculty of Science and Engineering. Ritsumeikan University
Bibliographic Information
- Other Title
-
- 波高分析用増幅器のアンダーシュートの改善
Abstract
It is very important to improve the undershoot on the amplifier system for the progression of the accuracy on the pulse height analysis. One of the undershoot occurs on the connecting points of the multi-stage amplifier and the selection of the small output impedance and the large coupling condenser at one stage is able to diminish it. Since the another one due to the differential characteristic of the radiation detector appears on the output of the pulse shaper, it is possible to cancel it by the delaied output signal of the radiation detector. In this paper, the theoretical cancellation by the adder and the experimental data are reported.
Journal
-
- Oyo Buturi
-
Oyo Buturi 36 (10), 789-796, 1967
The Japan Society of Applied Physics
- Tweet
Details 詳細情報について
-
- CRID
- 1390001204629704832
-
- NII Article ID
- 130003588455
-
- ISSN
- 21882290
- 03698009
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed