頭部CT検査における撮影角度及び範囲の自動設定の試み

書誌事項

タイトル別名
  • Automatic determination of Scan angle and range for Brain CT

この論文をさがす

説明

Reference lines of brain computed tomography (CT) is orbitomeatal (OM) line that runs through the external canthus and the center of the external auditory meatus. However, because it is difficult to identify OM line on the scout image which is 2-dimensional X-ray image, OM line is usually set by radiological technologists subjectively. The purpose of our study is to propose a method which detects the scan angle and the scan range used on the scout images automatically. The algorithm in our proposed method is based on a two-dimensional post-processing that is the method of Hough transformation conversion for detecting the scan angle and the Cross-correlation technique for the scan range. Our proposed method was tested using the scout images of 30 clinical patients' data by visual evaluation of four radiological technologists. Four radiological technologists evaluated the results of the clinical cases and 28 cases (93%) of CT examinations with using our method were acceptable. We confirmed it has the potential to improve the consistency in scan planning of brain CT, because the proposed method detects almost accurately the scan angle and the scan range in clinical cases. In conclusion, it is possible to obtain reproducible slice level images.

収録刊行物

被引用文献 (1)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ