Elimination of Interferences in the Mercury(II) Determination with a Dithizone Nanofiber-coated Test Strip Based on Filtration–Enrichment

Bibliographic Information

Other Title
  • ジチゾンナノ薄膜試験紙を用いる通液濃縮法による水銀イオン検出における妨害除去法の検討
  • ジチゾンナノ ハクマク シケンシ オ モチイル ツウエキ ノウシュクホウ ニ ヨル スイギン イオン ケンシュツ ニ オケル ボウガイ ジョキョホウ ノ ケントウ

Search this article

Abstract

This paper reports the selectivity of dithizone dye nanofiber-coated test strip (Dith DNTS) for the detection of HgII at ppb level. The Dith DNTS having thin layer (less than 500 nm) composed of 100% dithizone ensures highly sensitive HgII detection in the range of 0.1 to 20 ppb with the 3 σ detection limit of 0.057 ppb within 17 min by filtration enrichment of 100 mL of sample solution and subsequent reflection absorption measurement. The detection method has been shown to possess high tolerance to various cations and anions based on solid-liquid separation between dithizone layer and aqueous solutions in the presence of suitable reagents. Consequently, the allowable limits for reliable detection of 10 ppb of HgII with their removal requirements were as follows : Na+ (10000 ppm), K+ (5000 ppm), CaII (5000 ppm), AlIII (500 ppm), CrIII (200 ppm), CrVI (3 ppm), MnII (100 ppm), FeIII (100 ppm), NiII (100 ppm), CdII (10 ppm) and PbII (100 ppm) under acidic conditions, CuII (6.4 ppm) and ZnII (100 ppm) using EDTA as a masking reagent to form water-soluble chelates, FeII (70 ppm) with KBrO3 to oxidize FeII to FeIII, and Ag+ (50 ppb) by filtration a sample solution containing NaI through a membrane filter to remove AgI precipitates. This highly sensitive and selective detection system is a promising method for simple and reliable on-site monitoring of HgII in real water samples.

Journal

  • BUNSEKI KAGAKU

    BUNSEKI KAGAKU 61 (2), 123-126, 2012

    The Japan Society for Analytical Chemistry

Citations (2)*help

See more

References(24)*help

See more

Details 詳細情報について

Report a problem

Back to top