A New Automatic Calibration System for Polygon Mirrors
-
- MASUDA Tadashi
- 正会員 電気通信大学電気通信学部
-
- KAJITANI Makoto
- 正会員 電気通信大学電気通信学部
-
- TOYODA Kouji
- 正会員 計量研究所
Bibliographic Information
- Other Title
-
- ポリゴン鏡の校正の自動化に関する研究
- ポリゴンキョウ ノ コウセイ ノ ジドウカ ニカンスルケンキュウ
Search this article
Description
This paper describes a new automatic calibration system for polygon mirrors. This system consists of the following four subsystems: (a) a precise rotary mechanism, (b) a rotary magnetic scale for angular standard, (c) a special photo-electric autocollimator which enables a dynamic measurement and (d) a mini-computer system which controls the compiling and handling of data. The basic principle of the system depends on the selfchecking-method, so that both the polygon mirror and the standard rotary scale are calibrated simultaneously. The system adopts a time-conversionmethod to detect slight angular difference with high resolution. By way of an example, a 12-sidedpolygon has been calibrated on the system. The results obtained are as follows: (1) The calibration can be done automatically in a short time without skill, excepting the manual operation of setting a polygon relatively for the magnetic scale. (2) 959% confidence interval of instrumental error calibrated on the system is within ±0.07 seconds of arc.
Journal
-
- Journal of the Japan Society of Precision Engineering
-
Journal of the Japan Society of Precision Engineering 49 (7), 904-909, 1983
The Japan Society for Precision Engineering
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001204732074240
-
- NII Article ID
- 130003784740
- 110001291655
-
- NII Book ID
- AN00130014
-
- NDL BIB ID
- 2598241
-
- ISSN
- 03743543
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed