Factors Affecting Arterial Blood Pressure Monitoring

  • FUJIWARA Shigeki
    Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University
  • TACHIHARA Keiichi
    Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University
  • KAWAKUBO Yoshifumi
    Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University
  • MORI Satoshi
    Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University
  • TOYOGUCHI Izumi
    Argon Medical Devices Japan K.K.
  • YOKOYAMA Takeshi
    Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University

Bibliographic Information

Other Title
  • 動脈血圧モニタリングキットに影響を与える因子について

Search this article

Description

Pressure transducer monitoring kits have their own frequency characteristics (natural frequency, damping coefficient). These frequency characteristics are influenced by many factors. A PlanectaTM (JMS, Hiroshima, Japan) for blood sampling is one factor that has a remarkable influence on the natural frequency in the pressure transducer monitoring kits. Resonance Over-Shoot Eliminator (ROSETM) has been released by Argon Medical Devices (TX, USA). This device increases damping coefficient by suppressing resonance phenomenon in the pressure transducer monitoring kit. We describe the influence of the insertion of one or two PlanectasTM and insertion of the ROSETM on frequency characteristics.