- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Influence of Doppler Angle in Carotid Artery Duplex Ultrasound Evaluation
-
- Sugawara Masayo
- Vascular Board, Department of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan
-
- Akai Atsushi
- Department of Vascular Surgery, The University of Tokyo Hospital, Tokyo, Japan
-
- Chigira Mayumi
- Vascular Board, Department of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan
-
- Okano Tomoko
- Vascular Board, Department of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan
-
- Okamoto Hiroyuki
- Department of Vascular Surgery, The University of Tokyo Hospital, Tokyo, Japan
-
- Shigematsu Kunihiro
- Department of Vascular Surgery, The University of Tokyo Hospital, Tokyo, Japan
-
- Miyata Tetsuro
- Vascular Center, Sanno Hospital and Sanno Medical Center, Tokyo, Japan
Bibliographic Information
- Other Title
-
- 超音波検査による内頸動脈狭窄診断におけるドプラ入射角の影響
Search this article
Description
We investigated influence of Doppler angle on peak systolic velocity (PSV). In our hospital, 109 carotid arteries were examined using carotid duplex. The PSVs obtained with a 45˚ and a 60˚ insonation angles at the distal common carotid artery and proximal internal carotid artery were statistically analyzed. The PSV measurements obtained with the two fixed insonation angles were statistically significantly different (p < 0.0001). Doppler insonation angle has a significant influence on PSV. Carotid duplex examination should be performed with a fixed angle in order to obtain reproducible results.
Journal
-
- The Journal of Japanese College of Angiology
-
The Journal of Japanese College of Angiology 54 (7), 91-96, 2014
Japanese College of Angiology
- Tweet
Details 詳細情報について
-
- CRID
- 1390001204903134208
-
- NII Article ID
- 130004567694
-
- ISSN
- 18808840
- 03871126
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed