A study on the hydrodynamics of a new bileaflet valve prosthesis using an aluminum alloy.

DOI
  • KIKUTA Y.
    Department of Biomedical Engineering School of Engineering University of Hokkaido
  • YUHTA T.
    Department of Biomedical Engineering School of Engineering University of Hokkaido
  • 阿部 潤
    北海道大学工学部生体機能合成講座
  • SHIMOOKA T.
    Department of Biomedical Engineering School of Engineering University of Hokkaido
  • MITAMURA Y.
    北海道東海大学工学部電子工学科

Bibliographic Information

Other Title
  • アルミニウム基金属を用いた中心開放型機械弁の流体動態的特性評価

Search this article

Abstract

In this paper we are reporting on a new bileaf let prosthetic heart valve made of an aluminum alloy. Test valves were made to observe characteristics of hydrodynamics around the valves, and to determine the basic valve design. In order to determine the optimal valve design, we had two design parameters which are axis position rate and leaflet curvature. Results of hydrodynamic studies suggested that the optimal combination between axis position rate and leaflet curvature were 70% and 9%, respectively among the test valves. The new bileaflet valve shows almost the same hydrodynamic characteristics as a conventional bileaflet valve.

Journal

  • Jinko Zoki

    Jinko Zoki 21 (2), 620-624, 1992

    JAPANESE SOCIETY FOR ARTIFICIAL ORGANS