人工弁による大動脈内乱流特性のレーザードップラ計測

DOI

書誌事項

タイトル別名
  • Laser Doppler measurement of turbulent characteristics in the aortic arch responding to the prosthetic valve orientations.

この論文をさがす

抄録

Local distributions of pulsatile flow velocity and turbulent intensity were measured in the model of human aorta in response to the valve orientation of BSS and SJM valves by using laser Doppler velocimeter. Total assessment of the turbulent characteristics, energy loss and performance index of the valves presented that the outer curve orientation of BSS and the orientation 2 of SJM setting the connmissure in parallel to the aortic curvature provided lower turbulent intensity and the best hydrodynamic performance.

収録刊行物

  • 人工臓器

    人工臓器 18 (2), 706-709, 1989

    一般社団法人 日本人工臓器学会