Development and Validation of an Analytical Method for Bromate by Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS)

Bibliographic Information

Other Title
  • 液体クロマトグラフィータンデム質量分析による水道水中の臭素酸イオン分析条件の検討と妥当性評価
  • エキタイ クロマトグラフィータンデム シツリョウ ブンセキ ニ ヨル スイドウ スイチュウ ノ シュウソサン イオン ブンセキ ジョウケン ノ ケントウ ト ダトウセイ ヒョウカ

Search this article

Description

We developed a liquid chromatography–tandem mass spectrometry (LC/MS/MS) analysis method for the determination of bromate in drinking water. Furthermore, we have conducted a validation test along with 23 organizations to confirm the validity of the analytical method. Recovery tests of bromate spiked into drinking water obtained by each organization were performed with two set points, that is the standard value (0.01 mg L-1) and 1/10 of the standard value (0.001 mg L-1) of bromate in the Japanese Waterworks Act. The linearity of calibration curves, selectivity, accuracy (recovery) , repeatability, and intermediate precision of bromate concentrations were calculated by the data of the recovery tests. The linearity of the calibration curves and selectivity were satisfactory. The accuracy, repeatability, and intermediate precision for the two concentrations obtained by the recovery tests satisfied the criteria in the guideline for the validation of the testing method for drinking water. These criteria are provided by the Ministry of Health, Labour and Welfare, Japan. Therefore, we concluded that the analytical method developed in the present study is valid for the analysis of bromate in drinking water.