Nonresonant transducers for ultrasound imaging.
-
- Lewin Peter A.
- Department of Electrical and Computer Engineering and Biological Engineering and Science Institute, Drexel University
-
- Zhang Qian
- Department of Electrical and Computer Engineering and Biological Engineering and Science Institute, Drexel University
書誌事項
- タイトル別名
-
- Nonresonant transducers for ultrasound
この論文をさがす
説明
A new class of super wideband, nonresonant and high efficiency ultrasound imaging transducers is described. The design developed uses PVDF piezoelectric material and exhibits pulse-echo sensitivity on a par with that achievable with a conventional, resonant design employing PZT or PZT composite material. The transducers can operate at virtually any clinically relevant frequency which minimizes the trade-off between the achievable penetration depth and desired image resolution and are well suited for both conventional medical imaging in the frequency range 2-12 MHz and high frequency imaging such as ultrasound backscatter microscopy using the frequencies between 20 and 100 MHz. Both simulation results and preliminary experimental data are presented and show very good agreement. The potential for using this approach as a technological breakthrough in the imaging array design is pointed out.
収録刊行物
-
- Journal of the Acoustical Society of Japan (E)
-
Journal of the Acoustical Society of Japan (E) 18 (5), 215-222, 1997
一般社団法人 日本音響学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001205088426752
-
- NII論文ID
- 110003105491
-
- NII書誌ID
- AA00256597
-
- ISSN
- 21853509
- 03882861
-
- NDL書誌ID
- 4287961
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDLサーチ
- Crossref
- NDLデジコレ(旧NII-ELS)
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可