メタノールをキャリヤー液とする2,3-ジヒドロキシナフタレンによる鉄鋼中ホウ素のフローインジェクション蛍光分析

書誌事項

タイトル別名
  • Flow Injection Fluorometric Determination of Boron in Steel with 2,3-Dihydroxynaphthalene Using Methanol as a Carrier Liquid
  • メタノール オ キャリヤーエキ ト スル 2 3 ジヒドロキシナフタレン ニ ヨル テッコウ チュウ ホウソ ノ フローインジェクション ケイ コウブンセキ

この論文をさがす

説明

The size of the blank value is enumerated in one of the factor that influence the detection limit in the fluorescence determination method. The blank value in this case is brought about by excess reagent. Therefore, the some methods for removing the excess reagent were studied under the conditions of the stability of boron-complex. To remove the influence of excess reagent, the solvent–extraction method etc. is used beside a method of changing the pH. This study attempted a fluorescence determination of boron by flow injection analysis using the methanol as a carrier liquid. As a result, the amount of reagent decreased down to 1/15 compared to that of conventional method, because of decreasing the amount of water in the sample solution.<br>The water in the sample solution suppressed the reaction between boron and 2,3-dihydroxynaphthalene (2,3-DHN). Therefore, the decrease of water tolerates the low concentration of 2,3-DHN. Under the optimum conditions, a calibration curve was drawn, which was linear over the range of 0 to 120 ppb. The relative standard deviation at 20 ppb B was 0.9% (n=6). For the determination of a trace amount of boron in steels, the on-line separation/preconcentration of boron from iron matrix (500 ppm) was performed by using resin (Amberlite IRA 743). The limit of B determination was 0.8 ppb (1.5 ppm in steels). The determination results for boron in standard steel materials showed good agreement with the certified values.

収録刊行物

  • 鉄と鋼

    鉄と鋼 97 (2), 65-69, 2011

    一般社団法人 日本鉄鋼協会

参考文献 (11)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ