A Biplanar System to Assess Alignment and Position of Artificial Knee Joint in 3D Using Slot Radiography and 3D Models
-
- KOBAYASHI Koichi
- 新潟大学医学部保健学科
-
- TAKAHASHI Takumi
- 新潟大学大学院自然科学研究科
-
- OKANIWA Daisuke
- 新潟大学工学部機械システム工学科
-
- SAKAMOTO Makoto
- 新潟大学医学部保健学科
-
- TANABE Yuji
- 新潟大学工学部機械システム工学科
-
- SATO Takashi
- 新潟医療センター整形外科
-
- OMORI Go
- 新潟医療福祉大学健康科学部健康スポーツ学科
-
- KOGA Yoshio
- 二王子温泉病院
Bibliographic Information
- Other Title
-
- 2方向スロットラジオグラフィと3次元形状モデルによる人工膝関節アライメントと人工関節設置位置評価法
- 2 ホウコウ スロットラジオグラフィ ト 3ジゲン ケイジョウ モデル ニ ヨル ジンコウ シツカンセツ アライメント ト ジンコウ カンセツ セッチ イチ ヒョウカホウ
Search this article
Abstract
The purpose of this study was to develop a method for evaluation of the three-dimensional (3D) alignment and position of the artificial knee joint using a biplanar slot radiography (SR) system and 3D models. A SR system uses collimated fan beam x-rays synchronized with the movement of a flat-panel detector. This system allows to obtain a full length x-ray image of the body with reduced dose and smaller image distortion compared with conventional x-ray systems. The biplanar SR system used a rotation table positioned at 0 deg and 60 deg relative to the optical axis of an X-ray source. Camera calibration was performed using an acrylic frame with 72 radiopaque markers. The femoral and tibial components of a total knee replacement system were immobilized and imaged using the biplanar SR system. Then, the femoral component was installed on a sawbone femur and imaged again. The 3D positions of the components and femur were recovered using an interactive 2D to 3D image registration method. Overall, the largest estimation errors were within 1 mm in translation and 1 deg in rotation, demonstrating that this method has an adequate accuracy for the clinical usage.
Journal
-
- Journal of the Japanese Society for Experimental Mechanics
-
Journal of the Japanese Society for Experimental Mechanics 16 (2), 134-139, 2016
The Japanese Society for Experimental Mechanics
- Tweet
Details 詳細情報について
-
- CRID
- 1390001205190827264
-
- NII Article ID
- 130005165137
-
- NII Book ID
- AA11822914
-
- ISSN
- 18844219
- 13464930
-
- NDL BIB ID
- 027556308
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- CiNii Articles
- KAKEN
-
- Abstract License Flag
- Disallowed