ヌル点クロスカップリングによるやや離れた同調回路の共振周波数の非接触測定とその体内計測への応用
書誌事項
- タイトル別名
-
- Null-point cross-coupling for semi-remote sensing of resonant frequency of tuning circuit and its application for intra-corporeal sensing
説明
<p>When two flat ring coils are layed on a plane in free space, overlapping about 20%, there can be found a null-coupling point where mutual flux coupling is cancelled to make zero transmission to each other. On such setup we can detect nearby interrupting magnetic material and/or LC-resonant tank circuit which causes extra transmission via cross-coupling to the interrupter. Such detection capabikity would depend on accuracy of null-point setup, driving signal design, gain and bandwidth of detecting amplifier and the size of the interrupter to be detected. Adequate design can detect a similar size tank circuit having Q0=70 or more at several times distance of interlogating coil pair diameter. In this study we aimed to use this for extrra corporeal sensing of intra corporeal passive sensing device. In our model experiment 5cm interlogating coil pair can steadily sense 1cm dia. x 1cm length ferrite cored LC tank circut, at 1 to 2 MHz frequency band.</p>
収録刊行物
-
- 生体医工学
-
生体医工学 54Annual (Proc), 1T5-4-5-1-1T5-4-5-2, 2016
公益社団法人 日本生体医工学会