Evaluation of ferroelectric hysteresis loops of leaky multiferroic BiFeO3 films using a system with a high driving frequency of 100 kHz system
-
- NAGANUMA Hiroshi
- Department of Applied Physics, Faculty of Science, Tokyo University of Science
-
- INOUE Yosuke
- Department of Applied Physics, Faculty of Science, Tokyo University of Science
-
- OKAMURA Soichiro
- Department of Applied Physics, Faculty of Science, Tokyo University of Science
この論文をさがす
抄録
A measurement system with a high driving frequency of 100 kHz is effective for measuring the ferroelectricity of leaky ferroelectric materials such as multiferroic BiFeO3 films. A high driving frequency reduces the measurement time, leading to a drastic reduction in the influence of the leakage current density on ferroelectric hysteresis loops. Phase-delay compensation is essential in a system with a driving frequency of 100 kHz; in this study a standard capacitor was used for phase-delay compensation. The value of remanent polarization of a BiFeO3 film measured by the 100-kHz system was confirmed by a positive, up, negative and down measurement.
収録刊行物
-
- Journal of the Ceramic Society of Japan
-
Journal of the Ceramic Society of Japan 118 (1380), 656-658, 2010
公益社団法人 日本セラミックス協会