Evaluation of ferroelectric hysteresis loops of leaky multiferroic BiFeO3 films using a system with a high driving frequency of 100 kHz system

  • NAGANUMA Hiroshi
    Department of Applied Physics, Faculty of Science, Tokyo University of Science
  • INOUE Yosuke
    Department of Applied Physics, Faculty of Science, Tokyo University of Science
  • OKAMURA Soichiro
    Department of Applied Physics, Faculty of Science, Tokyo University of Science

この論文をさがす

抄録

A measurement system with a high driving frequency of 100 kHz is effective for measuring the ferroelectricity of leaky ferroelectric materials such as multiferroic BiFeO3 films. A high driving frequency reduces the measurement time, leading to a drastic reduction in the influence of the leakage current density on ferroelectric hysteresis loops. Phase-delay compensation is essential in a system with a driving frequency of 100 kHz; in this study a standard capacitor was used for phase-delay compensation. The value of remanent polarization of a BiFeO3 film measured by the 100-kHz system was confirmed by a positive, up, negative and down measurement.