誘導型アプリケータによる収束深部加温

  • 加藤 博和
    岡山大学大学院保健学研究科放射線技術科学分野
  • 黒田 昌宏
    岡山大学大学院保健学研究科放射線技術科学分野
  • 澁谷 光一
    岡山大学大学院保健学研究科放射線技術科学分野
  • 金澤 右
    岡山大学大学院医歯薬学総合研究科病態制御科学専攻腫瘍制御学

書誌事項

タイトル別名
  • Focused Deep Heating with an Inductive Type Applicator

この論文をさがす

抄録

A new heating technique is proposed for deep heating using an annular-shaped Inductive Aperture-Type Applicator (IATA). It was shown using finite element analysis that this heating technique could concentrate RF energy in the central or core portion of the human body to generate heat without causing excessive heating of the outer or fat layers.<BR> A fat-muscle cylindrical phantom with a diameter of 0.30 m was inserted into the opening of an annular-shaped IATA, which has an outer diameter of either 0.74 or 0.94 m, a height of 0.20 or 0.30 m, and an opening 0.34 m in diameter, and is excited by RF currents of 13.56 or 27.12 MHz. The heat generated in the central portion of the phantom was 1.11-1.27 times that of the maximum value in the peripheral muscle portion of the phantom, while heat generated in the fat layer was only 20% of that generated in the central portion of the phantom.

収録刊行物

  • Thermal Medicine

    Thermal Medicine 23 (3), 133-143, 2007

    日本ハイパーサーミア学会

参考文献 (45)*注記

もっと見る

関連プロジェクト

もっと見る