Measurements of Characteristic Curves in a Computed Radiographic System (II)
-
- MORISHITA Junji
- Department of Radiology, Yamaguchi University Hospital, Ube City
-
- FUJITA Hiroshi
- Department of Electrical Engineering, Gifu National College of Technology
-
- SAKAMOTO Kiyoshi
- Department of Radiology, Hyogo College of edicine Hospital, Nishinomiya City
-
- UEDA Katsuhiko
- Department of Radiology, Yamaguchi University Hospital, Ube City
-
- OHTSUKA Akiyoshi
- Department of Radiology, Yamaguchi University Hospital, Ube City
-
- FUJIKAWA Tsuyoshi
- Department of Radiology, Onoda Cross Hospital, Onoda City
-
- YAMAUCHI Shuichi
- Department of Radiology, Yamaguchi University Hospital, Ube City
-
- KAMAKURA Toshiko
- Department of Radiology, Hyogo College of edicine Hospital, Nishinomiya City
-
- TANAKA Sadato
- Department of Radiology, Yamaguchi University Hospital, Ube City
Bibliographic Information
- Other Title
-
- コンピューテッドラジオグラフィの特性曲線の測定(II)
Description
We measured the “digital” characteristic curve, which relates the pixel value to the relative x-ray exposure, in a Fuji computed radiographic system (FCR-101). The time-scale sensitometric technique was applied to the sensitometry in the digital system due to the absence of the reciprocity-law-failure effect. The calibration curve showing the relationship between the relative x-ray exposure and the selected time in the x-ray generator was measured with a fluorescence meter and was used for determination of the relative x-ray exposure. It was found at two facilities that the digital characteristic curve obtained by the time-scale method was in good agreement with that obtained by inverse-square method. Moreover, the curves measured with different imaging parameter settings of the FCR system were comparable at two facilities. It was shown, at extremely larger value settings of imaging parameters, that the curve shape was nonlinear at higher and/or lower pixel values due mainly to the property of the electronic component such as photomultiplier. The use of the time-scale technique in the characteristic curve determination makes the measurements easier and shorter
Journal
-
- Medical Imaging and Information Sciences
-
Medical Imaging and Information Sciences 6 (1), 25-38, 1989