Micro Scan WalkAway POSシリーズパネルにおける<i>S. aureus</i>のVCM 2 μg/m<i>l</i>のデータに関する検討
書誌事項
- タイトル別名
-
- Study of VCM 2 μg/m<i>l </i>of <i>S. aureus </i>in the Micro Scan Walk Away POS series panel
この論文をさがす
説明
Using 71 Staphylococcus aureus isolates that are known to have a minimum inhibitory concentration (MIC) of 2 μg/ml for vancomycin (VCM), we compared the MICs obtained when the MicroScan Prompt system and manual preparation methods were used for preparing the bacterial solutions. For this analysis, the inocula were adjusted to match a 0.5 McFarland turbidity standard. We also compared the MICs seen in the case of the Pos Combo 3.1 J (PC 3.1 J) and Pos MIC 3.3J (PM 3.3 J) panels. When the Prompt system was used, 68 and 46 isolates showed a VCM MIC of 2 μg/ml with the PM 3.3 J and PC 3.1 J panels, respectively. When the manual preparation method was used, 53 isolates showed a VCM MIC of 2 μg/ml with the PM 3.3 J panel and only 22 isolates showed this MIC with the PC 3.1 J panel. The remaining isolates had an MIC of 1 μg/ml. Because a VCM MIC of 2 μg/ml or higher is clinically important in patient management, it is necessary to determine the differences between the results for the 2 panels and the 2 methods of inoculum preparation.
収録刊行物
-
- 医学検査
-
医学検査 63 (1), 90-93, 2014
一般社団法人 日本臨床衛生検査技師会