Evaluation of the automated hematology analyzer XN-550 for cerebrospinal fluid cell count
-
- TANAKA Masami
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- SHUKUYA Ken-ichi
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- MORITA Yoshifumi
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- HISASUE Naoko
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- KAGEYAMA Yuko
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- OKUBO Shigeo
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- SHIMOSAWA Tatsuo
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- YATOMI Yutaka
- Department of Clinical Laboratory, The University of Tokyo Hospital
Bibliographic Information
- Other Title
-
- 多項目自動血球分析装置XN-550を用いた髄液細胞数算定の基礎的検討
Search this article
Abstract
We evaluated the clinical usefulness of XN-550, an automated hematology analyzer (Sysmex, Japan), for the determination of cell count in cerebrospinal fluid samples. Reproducibility, linearity, sensitivity and correlation of findings with those of conventional examination using a microscope were good and reliable. Drainage samples, including deformed cells, were analyzed. However, the scattergram did not correspond to the microscopy examination results, and the number of deformed cells was classified as low among side scatter cells. XN-550 is easy to handle and provides quick results, and is therefore useful for screening examination.
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 64 (6), 749-754, 2015
Japanese Association of Medical Technologists
- Tweet
Details 詳細情報について
-
- CRID
- 1390001205742627328
-
- NII Article ID
- 130005119255
-
- ISSN
- 21885346
- 09158669
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed