Radioimmunoassayを用いた血中1,25-Dihydroxyvitamin D測定キットの検討 : 各種ビタミンD代謝物との交差反応を中心に
書誌事項
- タイトル別名
-
- Two Radioimmunoassay Systems for Serum 1,25-Dihydroxy Vitamin D : Cross-reactivities of Various Vitamin D Metabolites
- Radioimmunoassay オ モチイタ ケッチュウ 1 , 25 Di
この論文をさがす
抄録
Recently, two radioimmunoassay systems with ^<125>I-labelled tracer for 1α,25-dihydroxyvitamin D(1,25(OH)_2D) have been developed: IDS-RIA and INCSTAR-RIA. These assay systems provide simple and rapid methods that do not require sample pre-purification by HPLC before the binding assay, and are shown to be useful especially in individuals who do not receive vitamin D treatment. However, its clinical use has been still limited because of the limited number of studies, which demonstrate the cross-reactivities of vitamin D metabolites and the bias from the administered vitamin D related medicine. In this study, we investigated the cross-reactivities of various vitamin D metabolites by adding those compound to normal pool serum and compared the assays with an accepted radioreceptor assay (RRA) in patients with end-stage renal diseases under 1α-OH-D_3 therapy. Regarding the cross-reactivity of 1,25(OH)_2D_3 as 100 %, the indexes of 23(S)25(R)-1,25(OH)_2D_3-26,23-lactone, 1,24,25(OH)_3D_3, 1,24(OH)_2D_3 and 1,25(OH)_2D_3-24-oxo-D_3 were 33.5%, 169%, 81.7% and 150% in IDS-RIA and 0%, 28.1%, 44.3% and 24.0% in INCSTAR-RIA, respectively. In patients with end-stage renal diseases, the values from each RIA were well correlated with those from RRA. In conclusion, these new RIA assay systems for serum 1,25(OH)_2D are useful methods in patients under the vitamin D treatment.
収録刊行物
-
- ビタミン
-
ビタミン 72 (11), 607-611, 1998
公益社団法人 日本ビタミン学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001205760441344
-
- NII論文ID
- 110002843452
-
- NII書誌ID
- AN00207833
-
- ISSN
- 2424080X
- 0006386X
-
- NDL書誌ID
- 4608436
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可