Development of Automatic Analysis for Dynamic Renal Function Study Using Computed Tomographic (CT) Images

Bibliographic Information

Other Title
  • 腎動態シンチグラフィにおける Computed Tomography(CT)画像を用いた自動解析法の開発
  • ジンドウタイ シンチグラフィ ニ オケル Computed Tomography CT ガゾウ オ モチイタ ジドウ カイセキホウ ノ カイハツ

Search this article

Abstract

Determination of the region of interest (ROI) for dynamic renal function has been highly discordant among operators because of the dependence on factors such as the rate of injection of radioactive medicines, constitution, and renal function. To simplify this problem, we developed a computer algorithm that provides automatic analysis for both localization of the kidney and automatic determination of ROIs using computed tomographic (CT) images. The bilateral kidneys were extracted from enhanced CT images, and the CT pixel size was adjusted to dynamic renal function images. The template-matching technique was used for these images, and the kidney renal location was analyzed on additional functional images constructed by dynamic renal function images. As a result, we were able to obtain time-activity curves of both renal function and quantified glomerular filtration rate (GFR). In conclusion, the computer algorithm we developed was considered to provide reliable results, apart from the variability among operators, because of its good reproducibility. (Article in Japanese)

Journal

References(5)*help

See more

Details 詳細情報について

Report a problem