Computed Radiography システムの入出力特性を取得するための改良型マルチスリット装置の製作–ヒール効果の影響の補正と高管電圧実験への適用–
書誌事項
- タイトル別名
-
- Fabrication of Improved Multi-slit Equipment to Obtain the Input-output Characteristics of Computed Radiography Systems: Correction of the Heel Effect, and Application to High Tube-voltage Experiments
- Computed Radiography システム ノ ニュウシュツリョク トクセイ オ シュトク スル タメ ノ カイリョウガタ マルチスリット ソウチ ノ セイサク : ヒール コウカ ノ エイキョウ ノ ホセイ ト コウカン デンアツ ジッケン エ ノ テキヨウ
この論文をさがす
抄録
Multi-slit equipment is a new experimental apparatus that can measure the input-output characteristics of a CR (computed radiography) system with limited influence of the fading effect. Kimoto et al. recently proposed a new type of multi-slit apparatus in which the multi-slit setup, the insertion region of the phosphor plate, and plate shielding are integrated to create a single handy-type item (an all-in-one type multi-slit apparatus). However, some problems remained unsolved. The aims of this study were to devise a setup for application to high tube voltage conditions, and to improve the all-in-one type multi-slit equipment so as to correct the heel effect. We examined the capabilities of our improved multi-slit equipment using diagnostic X-ray apparatus and found that it can obtain input-output characteristics with 5% accuracy for tube voltages of 40–140 kV and SID (source to image receptor distances) of 50–200 cm.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 70 (9), 867-876, 2014
公益社団法人 日本放射線技術学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001206361522816
-
- NII論文ID
- 130004691174
-
- NII書誌ID
- AN00197784
-
- COI
- 1:STN:280:DC%2BC2M7jsFOksg%3D%3D
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 025840613
-
- PubMed
- 25242595
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可