Comparison of Dose Distribution on Radiographic Film and Radiochromic Film for Intensity Modulated Radiation Therapy
-
- Sasaki Motoharu
- Department of Radiology, Tokushima University Hospital Institute of Health Biosciences The University of Tokushima Graduated School
-
- Tominaga Masahide
- Institute of Health Biosciences The University of Tokushima Graduated School
-
- Ikushima Hitoshi
- Institute of Health Biosciences The University of Tokushima Graduated School
-
- Kishi Taro
- Department of Radiology, Tokushima University Hospital
-
- Kawashita Tetsuya
- Department of Radiology, Tokushima University Hospital
-
- Hara Yasuo
- Department of Radiology, Tokushima University Hospital
-
- Fukunaga Yukiko
- Department of Radiology, Tokushima University Hospital
-
- Yamada Kenji
- Department of Radiology, Tokushima University Hospital
-
- Oita Masataka
- Graduate School of Health Science, University of Okayama
-
- Nishitani Hiromu
- Institute of Health Biosciences The University of Tokushima Graduated School
Bibliographic Information
- Other Title
-
- 強度変調放射線治療におけるradiographic filmと radiochromic filmの比較評価
- 臨床技術 強度変調放射線治療におけるradiographic filmとradiochromic filmの比較評価
- リンショウ ギジュツ キョウド ヘンチョウ ホウシャセン チリョウ ニ オケル radiographic film ト radiochromic film ノ ヒカク ヒョウカ
Search this article
Description
Purpose: Radiographic film is generally used for inspection of dose distribution in intensity modulated radiation therapy (IMRT) at many institutions. However, the distribution of filmless systems can be expected to be used increasingly in the future. Therefore, we confirmed the utility of radiochromic film by comparing it with radiographic film that does not need an automatic processor. Result: Difference in does measured by radiographic film and radiochromic film tended to increase in the low does area, but it was limited in a range of 1.5%. Conclusion: When the dose distribution was verified in a highly accurate radiation therapy such as IMRT, the results suggested that radiochromic film can be useful in addition to radiographic film.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 66 (5), 525-534, 2010
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001206361620480
-
- NII Article ID
- 10026453315
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 10721939
-
- PubMed
- 20628221
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed