1.5 T MR 装置における条件付きMRI 対応人工内耳の安全性に関する検証

書誌事項

タイトル別名
  • The Safety of MR Conditional Cochlear Implant at 1.5 Tesla Magnetic Resonance Imaging System
  • 1.5T MR ソウチ ニ オケル ジョウケン ツキ MRI タイオウ ジンコウ ナイジ ノ アンゼンセイ ニ カンスル ケンショウ

この論文をさがす

抄録

<p>In magnetic resonance imaging (MRI) examination of the patients with the cochlear implant, only limited data have a mention for safety information in the instruction manual supplied by the manufacturers. Therefore, imaging operators require more detailed safety information for implant device. We conducted detailed examination about displacement force, torque, and demagnetizing of the cochlear implant magnet based on American Society for Testing and Materials (ASTM) standard using the PULSAR and CONCERTO (MED-EL) with 1.5 tesla MRI system. As a result, the displacement force and the torque of the implant magnet were less than the numerical values descried in the manual. Therefore, these have almost no effect on the body under the condition described in a manual. In addition, the demagnetizing factor of the cochlear implant magnet occurred by a change magnetic field. The demagnetization depended on the direction of a line of magnetic force of the static magnetic field and the implant magnet. In conclusion, the operator must warn the position of the patients on inducing in the magnet room.</p>