差分マップ法および連続撮像法によるParallel MRI画像のSNR測定

書誌事項

タイトル別名
  • Signal-to-noise Ratio Measurement in Parallel MRI with Subtraction Mapping and Consecutive Methods
  • サブン マップホウ オヨビ レンゾク サツゾウホウ ニ ヨル Parallel MRI ガゾウ ノ SNR ソクテイ

この論文をさがす

説明

When measuring the signal-to-noise ratio (SNR) of an image the used parallel magnetic resonance imaging, it was confirmed that there was a problem in the application of past SNR measurement. With the method of measuring the noise from the background signal, SNR with parallel imaging was higher than that without parallel imaging. In the subtraction method (NEMA standard), which sets a wide region of interest, the white noise was not evaluated correctly although SNR was close to the theoretical value. We proposed two techniques because SNR in parallel imaging was not uniform according to inhomogeneity of the coil sensitivity distribution and geometry factor. Using the first method (subtraction mapping), two images were scanned with identical parameters. The SNR in each pixel divided the running mean (7 by 7 pixels in neighborhood) by standard deviation/√2 in the same region of interest. Using the second (consecutive) method, more than fifty consecutive scans of the uniform phantom were obtained with identical scan parameters. Then the SNR was calculated from the ratio of mean signal intensity to the standard deviation in each pixel on a series of images. Moreover, geometry factors were calculated from SNRs with and without parallel imaging. The SNR and geometry factor using parallel imaging in the subtraction mapping method agreed with those of the consecutive method. Both methods make it possible to obtain a more detailed determination of SNR in parallel imaging and to calculate the geometry factor.

収録刊行物

被引用文献 (18)*注記

もっと見る

参考文献 (30)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ