カテーテルアブレーションとPercutaneous Coronary Intervention(PCI)における被ばく低減対策の有効性の検証
書誌事項
- タイトル別名
-
- Inspection of Dose Reduction in Patients during Percutaneous Coronary Intervention and Catheter Ablation Procedures
- カテーテルアブレーション ト Percutaneous Coronary Intervention(PCI)ニ オケル ヒバクテイゲン タイサク ノ ユウコウセイ ノ ケンショウ
この論文をさがす
説明
With an increasing number of interventional radiology (IVR) procedures, it is a critical issue to control and reduce the radiation dose for patients by radiological technologists. In our study, we analyzed the usefulness of a provision for radiation reduction on catheter ablation and percutaneous coronary intervention (PCI) procedures based on the data from radiation information system (RIS). With regard to catheter ablation, 50% reduction was enabled with decreasing fluoroscopic and radiographic conditions regardless of each technique. Radiation reduction enabled a decrease in the fluoroscopic dose during PCI procedure. However, note that excessive radiation reduction does not show positive results of the radiation dose reduction. Moreover it leads to an increase in fluoroscopic time.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 68 (9), 1221-1230, 2012
公益社団法人 日本放射線技術学会