Optimal Beam Quality for Chest Digital Radiography
-
- Oda Nobuhiro
- Department of Radiological Technology, Faculty of Medical Science, Kyoto College of Medical Science
-
- Tabata Yoshito
- Department of Radiological Technology, Faculty of Medical Science, Kyoto College of Medical Science
-
- Nakano Tsutomu
- Previous address: Department of Radiology, University of Occupational and Environmental Health, Hospital
Bibliographic Information
- Other Title
-
- 胸部CR撮影のための適正な線質の検討
- キョウブ CR サツエイ ノ タメ ノ テキセイ ナ センシツ ノ ケントウ
Search this article
Abstract
To investigate the optimal beam quality for chest computed radiography (CR), we measured the radiographic contrast and evaluated the image quality of chest CR using various X-ray tube voltages. The contrast between lung and rib or heart increased on CR images obtained by lowering the tube voltage from 140 to 60 kV, but the degree of increase was less. Scattered radiation was reduced on CR images with a lower tube voltage. The Wiener spectrum of CR images with a low tube voltage showed a low value under identical conditions of amount of light stimulated emission. The quality of chest CR images obtained using a lower tube voltage (80 kV and 100 kV) was evaluated as being superior to those obtained with a higher tube voltage (120 kV and 140 kV). Considering the problem of tube loading and exposure in clinical applications, a tube voltage of 90 to 100 kV (0.1 mm copper filter backed by 0.5 mm aluminum) is recommended for chest CR.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 70 (11), 1265-1272, 2014
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001206362481664
-
- NII Article ID
- 130004713500
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 025935564
-
- PubMed
- 25410333
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
-
- Abstract License Flag
- Disallowed