低管電流撮影・フル再構成による前向き心電図同期320列面検出器CT冠動脈血管造影

書誌事項

タイトル別名
  • Prospective Electrocardiogram Gated Coronary 320-row Area Detector Computed Tomography Angiography Using Low Tube Current Scanning with Full Reconstruction
  • テイカン デンリュウ サツエイ ・ フル サイコウセイ ニ ヨル マエムキ シンデンズ ドウキ 320レツメン ケンシュツキ CT カンドウミャク ケッカンゾウエイ

この論文をさがす

抄録

Background: It is possible to obtain equivalent image quality and a lower radiation dose using low tube current scanning with full reconstruction as compared to usual tube current scanning with half reconstruction in a 320-row area detector computed tomography (ADCT) angiography. Method: Of 589 patient underwent coronary CT angiography (CCTA), 11 patients with (RR-PQ) ≥1069 ms were enrolled. In those patients, low tube current (50% mA) scanning with full or half reconstruction were performed. As a control, 11 patients with matched pairs of tube voltage, BMI and heart rate who underwent usual scanning with half reconstruction (100% mA with half reconstruction) were selected. Standard deviation of the CT value (SD) was measured in aorta (Ao), left atrium (LA) and left ventricle (LV), and extended dose-length products (DLP.e) were calculated. Result: Significant motion artifact was not observed in any patients. SD of 50% mA with half reconstruction, 50% mA with full reconstruction, and 100% mA with half reconstruction were 28.1±2.6, 20.3±1.9, 20.7±2.5 HU in Ao, 34.4±4.4, 24.9±2.8, 24.9±3.1 HU in LA, and 29.7±2.3, 21.7±1.9, 22.1±2.3 HU in LV, respectively. There were not significant differences between 50% mA with full reconstruction and 100% mA with half reconstruction, but there were significant differences between 50% mA with half reconstruction and 50% mA with full reconstruction in all sites. The DLP.e of 50% mA scanning (74.1±21.8 mGy·cm) was significantly lower than 100% mA scanning (161.9±28.9 mGy·cm). Conclusion: CCTA with lower radiation dose and equivalent image quality can be obtained by ADCT using 50% mA scanning with full reconstruction in patients with (RR-PQ) ≥1069 ms.

収録刊行物

参考文献 (16)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ