Quality Assessment of X-ray Images Using Modulation Transfer Function Fractal Dimension

  • Endo Shun
    Graduate School of Health Sciences, Niigata University
  • Lee Yongbum
    Graduate School of Health Sciences, Niigata University
  • Tsai Du-Yih
    Graduate School of Health Sciences, Niigata University
  • Hara Takeshi
    Graduate School of Medicine, Gifu University

Bibliographic Information

Other Title
  • MTFフラクタル次元によるX線画像の画質評価
  • MTF フラクタルジゲン ニ ヨル X センガゾウ ノ ガシツ ヒョウカ

Search this article

Abstract

Objective evaluation, e.g., using the modulation transfer function (MTF) or the noise power spectrum (NPS) of an X-ray imaging system may not correctly correspond to the subject’s evaluation, e.g., when using the receiver operating characteristic (ROC). Moreover, it is difficult to measure MTF or NPS from clinical images. We therefore applied MTF fractal dimension to an X-ray imaging system. The MTF fractal dimension includes the frequency properties of the human eye in addition to quantitative complexity. In this study, we demonstrated a relationship between basic image quality and MTF fractal dimension using simulated and actual X-ray images.

Journal

References(1)*help

See more

Details 詳細情報について

Report a problem