医用液晶ディスプレイを用いたX線画像観察による眼の疲労度の客観的な評価

書誌事項

タイトル別名
  • Objective Evaluation of Visual Fatigue for Reading of Radiographs Displayed on Medical-grade Liquid-crystal Displays
  • イヨウ エキショウ ディスプレイ オ モチイタ X センガゾウ カンサツ ニ ヨル メ ノ ヒロウド ノ キャッカンテキ ナ ヒョウカ

この論文をさがす

抄録

The aim of this study was to evaluate visual fatigue objectively by measuring accommodation time and critical fusion frequency (CFF) before and after reading posteroanterior chest radiographs displayed on medical-grade liquid-crystal displays (LCDs) under different monitor conditions. A color LCD (500, 170 cd/m2) and a monochrome LCD (500 cd/m2) were used in this study. Six observers independently kept reading the radiographs for two hours to understand various lung nodules in the “Fatigue Session”. Objective visual fatigue was measured by using the accommodation device and the CFF meter before and after the Fatigue Session. The ambient lighting of the laboratory was set at 35 lux. Both the accommodation time and the CFF between before and after the Fatigue Session indicated statistically significant differences (p<0.05). Our results on accommodation time and CFF before and after reading the radiographs on medical-grade LCDs indicated that visual fatigue could be evaluated objectively.

収録刊行物

被引用文献 (2)*注記

もっと見る

参考文献 (54)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ