AN EXPERIMENTAL STUDY OF THE MEDICAL 15MeV BETATRON : VARIATION OF THE DOSE RATE WITH DISTANCE
-
- MITSUDA HIDEO
- Department of Radiology, Osaka City University Medical School
-
- ITO KEIZO
- Department of Radiology, Osaka City University Medical School
Bibliographic Information
- Other Title
-
- 医用15MeV Betatronに関する実験的研究 : 距離と線量率の関係
- イヨウ 15MeV Betatron ニ カンスル ジッケンテキ ケンキュウ キョリ ト センリョウリツ ノ カンケイ
Search this article
Description
Variations dose rates of X-rays and electron beams of the medical 15 MeV betatron with forcus-ionization chaber distance are studied. As for X-rays, the low energy component is found to be fairly large, especially near the donut. Therefore the dose rate is not inversely proportional to the square of the distance as is simply expected. But using 2.5cm thick acrylite or 3mm lead cap on the ionization chamber, the inversely proportional relation can be obtained. The effect of the scattering rays from the filter is also examined by this method. In the case of the electron beams, the dose rate is not inversely proportional to the square of distance at all, but to the m-th power of the distance (2<m<3). The value of m differs for various energies or methods of measurement. The location of the electron beam source, so called the equivalent focus, is estimated from the variation of the dose rate with distance. The equivalent focus almost coincides with the center of the donut. But the ones decided by photographic method are not same point on the orbital plane and on the vartical plane. And moreover they vary with energy.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 18 (4), 270-274, 1963
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001206364088576
-
- NII Article ID
- 110003452686
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 9174753
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed