EFFECT OF THE ABSORBERS FOR THE DOSE DISTRIBUTION IN CONFORMATION RADIOTHERAPY OF CARCINOMA OF THE MAXILLARY SINUS

  • AOYAMA K.
    Department of Radiotherapy, Aichi Cancer Center Hospital
  • MUROGA M.
    Department of Radiotherapy, Aichi Cancer Center Hospital
  • UEDA T.
    Department of Radiotherapy, Aichi Cancer Center Hospital

Bibliographic Information

Other Title
  • 上顎癌原体照射時の眼球保護に使用される打抜き吸収体の材質の検討
  • ウワアゴガン ゲンタイ ショウシャジ ノ ガンキュウ ホゴ ニ シヨウ サレル ウチヌキ キュウシュウタイ ノ ザイシツ ノ ケントウ

Search this article

Abstract

In conformation radiotherapy combined with the hollow-out technique, it is essential to select the suitable material of the absorber, as the material of the absorber influences on the dose distribution. Practically, it is necessary to shield both-sided eyes, in the treatment of carcinoma of the maxillary sinus. The distortion of the dose distribution caused with the absorption of tungsten, copper and zinc were measured. In order to shield the healthy-sided eye completely, it was suitable to use tungsten rod (2cm. in diameter) as the absorber. With the use of tungsten rod to shield the ill-sided eye, its shielding effect was excellent, while the distortion of the dose distribution could not be ignored. By using the zinc rod, the distortion of the dose distribution deviated from the expected beam focus was neglisible, on the other hand the shielding to the ill-sided eye was considerably good.

Journal

Keywords

Details 詳細情報について

Report a problem

Back to top