Research on Time-course Differential Imaging by Bone Scintigraphy
-
- KAWAMURA SEIJI
- Center for Diagnostic Imaging, Kurume University Hospital
-
- FUKUSHIMA SHIGEHIRO
- Applied Systems Design Unit, Department of Applied Information and Communication Sciences, Faculty of Design, Kyushu University
-
- YOSHINAGA YUKIYASU
- Applied Systems Design Unit, Department of Applied Information and Communication Sciences, Faculty of Design, Kyushu University
-
- KAWATA HIDEMICHI
- Center for Diagnostic Imaging, Kurume University Hospital
-
- MIURA HISAO
- Center for Diagnostic Imaging, Kurume University Hospital
-
- MIYAGAWA TERUO
- Center for Diagnostic Imaging, Kurume University Hospital
-
- FUKUDOME YOSHIFUMI
- Center for Diagnostic Imaging, Kurume University Hospital
-
- UMEZAKI NORIYOSHI
- Department of Radiochemistry and Radiation Hygiene, Daiichi College of Pharmaceutical Sciences
-
- ISHIBASHI MASATOSHI
- Department of Radiology, Kurume University School of Medicine
-
- MORITA SEIICHIRO
- Department of Radiology, Kurume University School of Medicine
-
- HAYABUCHI NAOFUMI
- Department of Radiology, Kurume University School of Medicine
Bibliographic Information
- Other Title
-
- 骨シンチグラフィへの経時的差分画像法適用に関する研究
- ホネ シンチグラフィ エ ノ ケイジテキ サブンガ ゾウホウ テキヨウ ニ カンスル ケンキュウ
Search this article
Abstract
The temporal image subtraction technique was applied to bone scintigraphy, using Photoshop(commercially available image processing software) and Morpher (public domain warping software). For the temporal subtraction images, 81 subtraction images (19 cases) were prepared by a method used to subtract the previous images from the current ones. Registration of the current and previous images was performed by manual operation using Photoshop, and warping was done using the warping function of Morpher. In addition, difference images prepared after correcting the distributions of radioactive isotopes of the current and previous images using the count of the pelvic region were also examined. Compared with manual operation, alignment of images by warping improved registration and reduced the generation of pseudo-images of subtraction images. The rate of identification of abnormal accumulation-enhanced regions and subjective evaluation by doctors was improved for warping more than for manual operation. Furthermore, abnormal hot regions, which are difficult to find in film images, could be found in three subtraction images. In addition, it was confirmed that abnormal hot regions become more visible in many cases by preparing subtraction images after correcting the count between images using the count of the pelvic region. Thus, it is suggested that the temporal image subtraction technique in bone scintigraphy enables more accurate observation of enhancement of or changes in abnormal hot regions, which will support diagnostic reading. It is considered that enhancement of or changes in abnormal hot regions will be more accurately understood through further detailed discussion in the future.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 61 (7), 997-1005, 2005
Japanese Society of Radiological Technology
- Tweet
Details 詳細情報について
-
- CRID
- 1390001206389275008
-
- NII Article ID
- 110003467538
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
- http://id.crossref.org/issn/03694305
-
- NDL BIB ID
- 7430618
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
- KAKEN
-
- Abstract License Flag
- Disallowed